Alarm&Display Device ,AID150
Alarm&Display Device ,AID150
AID Series alarm and display devices are remote display and audible and visible alarm devices. AID duplicates fault, alarm and operating message of monitoring devices. When fault occurs, AID gives audible and visible signal to medical staff and technical staff. Additionally, response value of AIM-M series monitoring device can be set by AID remotely. AID Series alarm and display device have three types: AID10, AID120 and AID150, which shall be arranged in operating room or nursing station so that it can be permanently monitored by medical staff.
Technical parameters of AID120 / 150
Parameter Type |
AID120 |
AID150 |
|
Auxiliary power supply |
Voltage |
DC 24V |
|
Consumption |
< 0.6W |
||
Display range of insulation resistance |
0~999kΩ |
—— |
|
Insulation alarming range |
50~999kΩ |
—— |
|
Transformer load rate display |
Percentage display |
—— |
|
Load current alarm setting |
14A、18A、22A、28A、35A、45A |
||
Temperature alarm setting range |
0~+200℃ |
||
Alarm method |
Sound-light alarm |
||
Alarm type |
Insulation failure, overload, overheat, equipment failure |
||
Communication mode |
RS485, Modbus-RTU |
||
Display mode |
LED digital display |
128×64 LCD display |